ENVIRONMENTAL CHAMBERS SPECIFICATIONS
|
|
|
|
|
MODEL NUMBER
|
651.05F-01
|
651.05F-02
|
651.06E-03
|
651.06E-04
|
651.10E-04
|
Load frame
|
Acumen 1
Acumen 3
|
Landmark
Bionix
Acumen 3
|
Landmark
|
Landmark
|
Landmark
|
|
|
|
|
|
|
Chamber part number
|
058-052-704
(US Plug)
058-052-703
(WW Plug)
|
058-052-702
(Acumen, US Plug)
058-052-701
(Acumen, WW Plug)
058-052-602 (w/table top legs,
US Plug)
058-052-601 (w/table top legs,
WW Plug)
|
100-639-950 (works with
494 FlexTest Controller)
100-639-951 (works with 494 Flex Test Controller)
(will be supplied with defrosting elements for the window)
|
100-639-952 (works with
494 FlexTest Controller)
100-639-953 (works with 494 Flex Test Controller)
(will be supplied with defrosting elements for the window)
|
100-639-954 (works with
494 FlexTest Controller)
100-639-955 (works with 494 Flex Test Controller)
(will be supplied with defrosting elements for the window)
|
Cart part number¶
|
100-294-319#
100-297-346
|
100-294-319#
100-297-346
|
100-160-646
100-213-156**
|
100-160-646
100-213-156**
|
100-160-646
100-213-156**
|
Test space (nominal)
|
216 mm W,
228 mm D,
228 mm H
(8.5 in W, 9 in D, 9 in H)
|
286 mm W,
305 mm D,
457 mm H
(11.25 in W, 12 in D, 18 in H)
|
356 mm W,
432 mm D,
559 mm H
(14 in W, 17 in D, 22 in H)
|
356 mm W,
432 mm D,
813 mm H
(14 in W, 17 in D, 32 in H)
|
356 mm W,
432 mm D,
660 mm H
(14 in W, 17 in D, 26 in H)
|
Removable “U-Plug” sections
|
Not included
|
Included
|
Included
|
Included
|
Included
|
Typical application
|
Material testing, including fatigue and fracture, tension, compression, flex/bend, etc.
|
Air temperature
performance*
|
|
|
|
|
|
Temperature Values
(min & max)
|
-151°C (-238°F) to
350° (662°F) |
-151°C (-238°F) to
350° (662°F) |
-129°C (-200°F) to
315° (600°F) |
-129°C (-200°F) to
315° (600°F) |
-129°C (-200°F) to
540°C (1000°F)
|
Heating |
Ambient to +350°C
(+660°F) in 45 minutes
|
Ambient to +350°C
(+660°F) in 45 minutes
|
Ambient to +315°C
(+600°F) in 30 minutes
|
Ambient to +315°C
(+600°F) in 30 minutes
|
Ambient to +540°C
(+1000°F) in 70 minutes
|
LN2 cooling |
Ambient to -151°C
(-238°F) in 45 minutes
|
Ambient to -151°C
(-238°F) in 45 minutes
|
Ambient to -129°C
(-200°F) in 26 minutes
|
Ambient to -129°C
(-200°F) in 26 minutes
|
Ambient to -129°C
(-200°F) in 26 minutes
|
Temperature
gradient †, ‡
|
±3°C or less |
±3°C or less |
±2°C (±5°F) |
±2°C (±5°F) |
±2°C (±5°F) from
-129° to 260°C
(-200° to 500°F)
±5°C (±10°F) from
-260° to 540°C
(-500° to 1000°F)
|
Temperature stability† |
±2°C (±5°F) |
±2°C (±5°F) |
±1°C (±3°F) |
±1°C (±3°F) |
±1°C (±3°F) |
Cooling requirements |
|
|
|
|
|
LN2 cooling inlet fitting
|
Male 1/2 in SAE 45° flare
90° angle tube fitting |
Male 1/2 in SAE 45° flare
90° angle tube fitting
|
Male 1/2 in SAE 45° flare
90° angle tube fitting
|
Male 1/2 in SAE 45° flare
90° angle tube fitting
|
Male 1/2 in SAE 45° flare
90° angle tube fitting
|
LN2 pressure
|
0.15 MPa (22 psi nominal)
0.21 MPa (30 psi max)
|
0.15 MPa (22 psi nominal)
0.21 MPa (30 psi max)
|
0.15 MPa (22 psi nominal)
0.21 MPa (30 psi max)
|
0.15 MPa (22 psi nominal)
0.21 MPa (30 psi max)
|
0.15 MPa (22 psi nominal)
0.21 MPa (30 psi max)
|
Heater circuit power
|
208/230 V AC
|
208/230 V AC
|
208/230 V AC
|
208/230 V AC
|
208/230 V AC
|
Requirements
|
50/60 Hz, single-phase;
at least 20 A circuit
|
50/60 Hz, single-phase;
at least 20 A circuit
|
50/60 Hz, single-phase;
at least 30 A circuit
|
50/60 Hz, single-phase;
at least 50 A circuit
|
50/60 Hz, single-phase;
at least 50 A circuit
|
Weight
|
82 kg (180 lbs)
|
108 kg (250 lbs)
|
102 kg (225 lbs)
|
159 kg (350 lbs)
|
136 kg (300 lbs)
|
Light
|
None
|
120 V bulb
(40 watt recommended)
|
115 V bulb
(40 watt recommended)
|
115 V bulb
(40 watt recommended)
|
External 115 V bulb
(40 watt recommended)
|
* Performance data derived with chamber empty and access holes blocked. Consult MTS Systems for temperature performance for testing specimens having high thermal mass.
|
** For Landmark 370.50 only.
|
***Typical maximum temperature for general materials testing applications. Many factors, including thermal mass of grips and fixtures, impact maximum operating temperature.
|
¶ Consult MTS for cart information specific to your configuration.
|
|
|
|
# Cart stand for table top load units.
|
|
|
|
|
† After 30 minute stabilization time.
|
|
|
|
|
‡ Due to large variety of possible test setups an air gradient is given. Specimen gradient is much less.
|
|
|