Your MTS Acumen Test System is designed to require minimal maintenance, and MTS Routine Maintenance provides the correct amount of care needed for optimal performance. Regularly scheduled maintenance performed by MTS field service engineers can ensure proper working condition and extend the operating life of the load frame.
On a routine maintenance visit, MTS field service engineers check the hardware, electronics and current condition of the system. We inspect the crosshead operation, actuator condition, transducer alignment, cables, internal power supply and more. We will also verify overall system performance. At the end of each routine maintenance visit, we’ll provide a report of all services performed, current equipment condition and any recommendations, if necessary.
MTS Acumen™ Electrodynamic Load Frames
Calendar Time Using 8 Hour Run Time Rate Per Day
Daily
Monthly
Annually
Running Time - Hours
8
160
2,000
Clean Work Area / Machine Surfaces
X
Check Frame Mounted Control Emergency Stop
X
Check Remote Emergency Stop (If Present)
X
Check Test Area Enclosure - Open Door Applies Low Velocity
X
Load Frame / Crosshead
Check Crosshead Locks For Proper Operation
MTS
Check Crosshead Lift Belt For Wear
MTS
Check / Adjust Crosshead Lift Belt Tension
MTS
Check / Clean Crosshead Lift Screws
MTS
Check / Clean Columns
MTS
Assure Crosshead Movement Is Smooth
MTS
Actuator
Check Cooling System Is Operational
MTS
Clean Exposed Cooling Surfaces
MTS
Assure Actuator Movement Is Smooth
MTS
Check Actuator Guidance Mechanism
MTS
Check Brake Operation
MTS
Check Anti-Rotate Mechanism
MTS
Check For Excessive Contamination
MTS
Electrical
Assure Displacement Transducer Is Clean
MTS
Verify Displacement Transducer Is Aligned
MTS
Confirm Reference Transducer Is Aligned
MTS
Verify Reference Transducer Center Position
MTS
Check Internal Cables For Abrasion
MTS
Check Internal Cable Connectors
MTS
Check Internal Grounding
MTS
Check Internal Power Supplies
MTS
Clean Electronic Devices
MTS
System
Check Frame Mounted Control Emergency Stop
MTS
Check Remote Emergency Stop (If Present)
MTS
Check System Detects Velocity Limit When Enabled
MTS
Check Test Area Enclosure - Open Door Applies Low Velocity
MTS
Check Frame Mounted Control - High Power Prohibit Functions
MTS
Verify Unlocked Crosshead Generates Interlock
MTS
Check System Cable Bundle Connectors
MTS
Check System Cable Bundle For Abrasions
MTS
Check Supplemental Ground Attached
MTS
Verify Tuning Parameters Correct For Stable Control Modes
MTS
Verify Load Frame Protection Limits Correct And Enabled